CLINICAL APPLICATIONS

Activity Monitoring

Clinical context
Physical activity has substantial health benefits in the population with cardiovascular diseases and reduces cardiovascular risk. The risk of coronary artery disease is lower in active individuals compared with sedentary individuals. Regular exercise improves arterial compliance, which is a determinant of systolic blood pressure and pulse pressure, and endothelium function in coronary arteries. Physical training also increases baro-reflex sensitivity and enhances parasymphathetic activity. Finally, exercise training has a positive effect on Quality of Life in patients with cardiac failure, due to improved skeletal muscle strength and better endothelial function in peripheral arteries.

Novacor offers a unique solution for recording patient physical activity simultaneously with ECG Holter and/or BP. It uses the Quattrolter recorder and the Activity module of HolterSoft Ultima, based on a powerful mathematical signal processing algorithm. Clinically validated (Wetzler et al., BPM, 2003), the solution is fully automatic, does not require any presetting linked to patient gait. The software computes the patient activity (lying / sitting / standing / walking / cycling) and the activity level linearly related to walking (or cycling) speed.

Featuring patient activity monitoring, the Quattrolter provides a detailed trackability of the patient
diary and therefore helps clinicians to interpret ABPM levels and ECG abnormalities - often associated with physical stress - or to follow-up their patients during cardiac rehabitation program.

Novacor 's scientific publications

Activity and position monitoring during ambulatory blood pressure and heart rate monitoring

Validation of a two-axis accelerometer for monitoring patient activity during blood pressure or ECG holter monitoring.

ACTIVITY AND POSITION MONITORING DURING AMBULATORY BLOOD PRESSURE AND HEART RATE MONITORING

P. Gosse(1), P. Guillo(2), M.L. Wetzler(3)

(1) Hopital Saint André, Bordeaux, France
(2) Clinique Saint Yves, Rennes, France
(3) Novacor, Rueil-Malmaison, France

Objectives
The main limitation in the interpretation of ambulatory BP monitoring is linked to high BP variability with activity and position changes. The analysis and quantification of such changes is thus an important objective, especially when assessing the effects of antihypertensive drugs.

Methods
We tested a new device (Quatrholter, Novacor France) allowing both BP and continuous ECG monitoring, equipped with a thorax position sensor and a two direction accelerometer fixed on the patient's thigh. These 2 sensors give continuous information on pt's position and on activity of the legs. 24h monitoring during daily routine was performed twice, 7 days apart, in 18 normal volunteers. BP was measured every 15 ¢ during day and night.

Results
The information from the 2 sensors allowed to classify the activity of the patients in 5 groups: Lying asleep, Lying awake, Sitting, Standing immobile, Walking or running (with information on walking speed). BP and HR were significantly influenced by these 5 conditions. The reproducibility of BP and HR measurements between the 2 recordings was analysed according to these 5 conditions. Results for SBP are summarised in table with mean values for the 2 recordings, standard deviation of differences (SDD) and coefficient of correlation (r). As expected, BP and HR measured when asleep or in sitting position were the most reproducible. BP measured when standing immobile or moving is much more variable from a recording to another.

 
Mean
SDD
r

24h average (mmHg)
Daytime
Night-time
Lying
Sitting
Standing
Lying asleep
Lying awake
Sitting
Standing immobile
Walking

105.7
109.8
99.1
100.2
110.5
111.8
98.6
104.4
110.5
109.1
116.9

4.9
5.3
5.2
61.
4.3
9.5
4.4
9.3
4.3
11.3
9.7

0.84
0.85
0.85
0.72
0.93
0.55
0.87
0.67
0.93
0.4
0.66

Conclusion
This new device allows a reliable and efficient analysis of the patient's positions and activities during BP and HR monitoring. Analysis of BP and HR according to these information allow to limit the influence of changes in activity on the interpretation of the results.

Abstract: 4C.1
Journal of Hypertension Volume 23, Supplement 2, June 2005, page S135

TOP OF PAGE

Validation of a two-axis accelerometer for monitoring patient activity during blood pressure or ECG holter monitoring.
 
Wetzler ML et al.

OBJECTIVES
The aim of the study was to evaluate the efficiency of a position/activity monitoring system based on a dual-axis accelerometer strapped to the subject's thigh and a position sensor located within a monitor placed on the subject's belt

METHODS
Twenty-six subjects wearing two monitors (one accelerometer on each thigh) were submitted to various activities and positions under the control of an observer. An analysis of each tracing was performed both manually by a technician and automatically by dedicated software before being compared with the information gathered during the study.

RESULTS
The accelerometer allowed accurate discrimination between the standing versus the sitting and lying positions. The sitting and lying positions were correctly detected by the built-in position sensor provided the unit was firmly attached. Walking was adequately detected by the accelerometer. The activity score was well correlated with treadmill speed. Changes in position and activity were detected with a mean error of less than 3 s.

CONCLUSIONS
The combination of an accelerometer placed on the subject's thigh and a position sensor located at the subject's waist appeared to be a suitable system for position/activity monitoring during ambulatory ECG and blood pressure monitoring.

Wetzler ML et al.: Validation of a two-axis accelerometer for monitoring patient activity during blood pressure or ECG holter monitoring.

Download the article (PDF)

2003, Blood Press Monit., 8:229-35.